Products And Services
Stent Cutter
 
 
System-2007 Stent Cutter : Specifications

 

 
Laser:   Note- Laser may be Pulsed Nd: YAG,   Q-Switched Nd: YAG,  FIBER,  or DISK, spec will very depending upon choice
 
Type   Pulsed Nd: YAG
Modes Transverse (TEM00) or Low Order Multimode
Pulse Energy 2 - 15mJ nominal
Pulse Width 50 - 5000 microseconds
Rep Rate   1 - 2000Hz
Stability   10% peak to peak
Cut Kerf Width   0.0008" - 0.006" (20 - 152 microns)
 
Precision Motion System
 
Linear Stage
Travel   X: 200mm (servo controlled); Y: 10mm (manual)
Working Feedrate 0.05 - 1.6 in/sec  depending on geometric complexity
Accuracy 0.0001 in/in (2.54 microns / 25.4mm)
Resolution 0.00001in (0.254 microns)
Rotary Stage
Accuracy   0.028°
Resolution 0.0072°
Example On 0.060° (1.524mm) OD tubing this yields a resolution of 0.0000037in (0.094 microns) with an accuracy of 0.000015in (0.381 microns)
 
Workstation
 
Tube Capacity   Tube Capacity: OD: 0.010" - 0.400" (0.254 micron - 10.15 micron)
Length: 40" (100cm) standard, up to 72" (180cm) option
Footprint Workstation: 72" x 30" (183cm x 76cm)
Console: 31" x 22" (79cm x 56cm)
Power Supply: 36" x 24" (92cm x 61cm)
Features Class I safety interlocked laser enclosure, Class IV available
Vibration isolated laser and motion system
Programmable automated tube feed
Video system and monitor
 
Utility Requirements, spec will very depending upon laser
 
Power   Workstation: 208/220 VAC, 50/60 Hz, 40 amps
Console: 120 VAC, 50/60 hz, 15 amps
Water 20 liters/min at 25psig, 20°C